CLINITEK Atlas (rack i carousel)

Savršenstvo urinoanalize

CLINITEK Atlas® automatizovani analizator urina je stoni uređak koji je dizajniran za visoku frekventnost testova urina, koristeći dokazanu Siemens-ovu tehnologiju u dry-pad analizama biohemije urina.
Format
Roll or dry chemistry strips affixed to plastic support
Maximum Tests On-board
490
On-board Stability
14 days
Tests Measured by Dry Pad
Leukocytes, Ketones, Protein, Glucose, Nitrite, Occult Blood, Urobilinogen, pH, Bilirubin, Color, Creatinine, Protein-Low
Tests by Physical Determinations
Specific Gravity by refractive index method; clarity by measurement of transmitted and scattered light
Dry Pad Test Measurements
Color change measured by reflectance photometry. Dual readings at reactive and reference wavelength
CLINITEK Atlas System with Carousel Sample Handler – General
Dimensions
Depth: 64.8 cm (25.5 in)
Width: 72.4 cm (28.5 in)
Height:66.0 cm (26.0 in)
Weight
(instrument base only)
59.0 kg (130lb.)
Weight
(sample handler only)
7.2 kg (15.8 lb.)
Power Requirements
120 V 50/60 Hz, 100 V 50/60 Hz, 220-240 V 50/60 Hz
Ambient Operating Temperature Rate
18°C to 30°C (64°F to 86° F)
Ambient Operating Humidity Rate
20% to 80% relative humidity
Compliance
UL, CSA, CE
Throughput
1 sample every 16 seconds
Printer
External
Sample Handling
Trays
Circular sample tray
Tray Capacity
50 samples (can be used as 40 = position tray dedicated spaces for calibrators, controls, STATs)
Sample Loading Capacity
1 tray up to 50 samples
Sample Tubes
Various depending on sample tray. Lipped tubes such as URIN-TEK or lipless up to 10.5 cm long
Sample Volume
Any volume from 2 mL to maximum fill of tube
Pipette
Precision impedance; ±1 µL accuracy
Rinse Cycle
Automatic pipette and SG well rinse between samples
Stat Sample Handling
Can be inserted during run; up to 2 samples
Bar Codes
Code 39, Interleaved 2 or 5, Codabar, Code 128
On-Board Computer
Memory
1000 patient records plus 200 calibration control records
Interfaces
ASTM or mode to emulate CLINITEK® 200+
Display
12 lines by 40 characters VGA display
Calibration
Method
Wet calibration with 1 or 2 point calibration for all reagents
Frequency
With every Reagent Pak change (or every 24 hours when multiple sample-lot Reagent Paks used within 24 hours)