ADVIA 1800

Pomera granice produktivnosti

ADVIA® 1800 Clinical Chemistry System pomera granice produktivnosti. Proširen meni omogućava integraciju biohemijskih testova, testove za zloupotrebe lekova i narkotika, terapeutskih nivoa lekova (TDM), i specifičnih proteina kao što su Cystatin C and CardioPhase® hsCRP.

ADVIA 1800 Clinical Chemistry System nudi:
Kapacitet do 1800 testova na sat. Brzina koju je skoro nemoguće izložiti do maksimuma.
200 osnovnih metaboličkih panela na sat.
Veliki kapacitet za reagense i opcioni dodatak za reagense kako bi se smanjio broj prekida rada.
Optimalna standardizacija reagensa i sistema za efikasnu upotrebu.
Rashladnu jedinicu za kontrole i reagense kako bi bila povećana stabilnost i produktivnost.
Spremna za automatizaciju – nije potreban dodatni hardver.

General Chemistry
Alanine Aminotransferase1
Alanine Aminotransferase (P5P)1
Albumin (BCG)1
Albumin (BCP)
Alkaline Phosphatase (AMP)1
Ammonia
Amylase
Aspartate Aminotransferase1
Aspartate Aminotransferase (P5P)1
Bilirubin, Direct
Bilirubin, Total
Calcium (Arsenazo)1
Calcium (CPC)1
Carbon Dioxide, Enzymatic1
Chloride (ISE)
Cholesterol1
Cholinesterase
Creatine Kinase
Creatinine1
Creatinine, Enzymatic
Fructosamine
Gamma-Glutamyltransferase (GGT)
Glucose Hexokinase1
Glucose Oxidase1
HDL Cholesterol
Iron
Lactate
Lactate Dehydrogenase (L-P)
Lactate Dehydrogenase (P-L)
LDL Cholesterol
Lipase
Magnesium1
Pancreatic Amylase
Phosphorus1
Potassium
Sodium
Total Iron Binding Capacity (TIBC)
Total Protein1
Total Protein (Urine/CSF)
Triglycerides1
Urea Nitrogen1
Uric Acid1

Specific Proteins
α1- Acid Glycoprotein
α-1 Antitrypsin
Antistreptolysin O
Apolipoprotein A1
Apolipoprotein B
β2 Microglobulin
CardioPhase® high sensitivity CRP
Complement C3
Complement C4
C-Reactive Protein (CRP)
CRP, Wide Range (wr-CRP)
Cystatin-C
Ferritin
Haptoglobin
Immunoglobulin A (IgA)
Immunoglobulin G (IgG)
Immunoglobulin M (IgM)
Lipoprotein(a)
Myoglobin
Prealbumin
Rheumatoid Factor (RF)
Transferrin
Diabetes
Hemoglobin A1c
Microalbumin

Drugs of Abuse/Toxicology
6-Acetylmorphine (6-AM)3
Acetaminophen
Amphetamines
Barbiturates (Serum)
Barbiturates
Benzodiazepines (Serum)
Benzodiazepines
Cannabinoids (THC)
Cocaine Metabolite
Creatinine Specimen Validity Test2
Ecstasy
Ethanol
Methadone
Methadone Metabolites, EDPP
Nitrite Specimen Validity Test2
Opiates
Oxidant Specimen Validity Test2
pH Specimen Validity Test2
Phencyclidine
Propoxyphene
Salicylate
Specific Gravity Specimen Validity Test2
Tricyclic Antidepressants

Therapeutic Drug Monitoring
Carbamazepine
Digoxin
Gentamicin
Lithium
Phenobarbital
Phenytoin
Theophylline
Tobramycin
Valproic Acid
Vancomycin

1 Available in concentrated version
2 For forensic toxicology use only
3 Syva® EMIT® application

Type of System Random, continuous access, batch, discrete processing
Throughput Rate Up to 1800 tests/hour: 1200 tests/hour colorimetric, 600 tests/hour ISE
Assay Capacity Onboard 55 including 3 ISE
Sample Handling
Sample Tubes 5 mL, 7 mL, 10 mL tubes, 2 mL sample cups
Sample Tray 84 samples, positive sample identification
Sample Rack Handler (optional) Universal 5-position rack, 425 total onboard capacity, continuous feed capability
Sample Integrity Check Qualitative check for hemolysis, lipemia, and icterus
STAT Sample Loading 84, not dedicated
Bar Codes Interleaved 2 of 5, Code 39, Code 128, Codabar (NW7)
Auto-repeat Automatic repeat testing from the retained prediluted sample or original sample
Auto-dilution Automatic dilution from retained prediluted sample
Auto-reflex Testing Automatic ability to perform 3 additional tests based on results of first test
Primary Probe Liquid Level Sensing, Crash Protection, Clot/Clog Detection, Short Sample Detection
Microvolume Technology
Automatic Sample Predilution Samples diluted 1:5 (30 μL sample + 120 μL saline generates up to 15 tests), retained for auto-repeat until results are available
Predilution Tray 120 dilution cuvettes
Original Sample Volume 2 to 30 μL; average of 2-3 μL per test
Average Reagent Volume 80-120 μL per test
Storage Capacity Onboard 25,200 tests average; Usage of concentrated reagents increases maximum capacity
Reaction Area
Reaction Tray 221 reusable plastic cuvettes
Cuvette Optical Path Length 10 mm
Reaction Bath Inert fluorocarbon oil circulation system, 37°C
Photometer 14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596, 658, 694, 751, 805, 845, and 884 nm)
Light Source 12 V, 50 W halogen lamp, cooled by forced water circulation
Assay Methods Endpoint, rate reaction, 2-point rate, multipoint homogeneous immunoassay
Reaction Times 3, 4, 5, 10, 15, and 21 minutes
Automatic Correction Serum blank, cell blank, measurement point change, sample volume change in reassay
Point Forwarding Automatically extends linearity over assay range samples
Reagent Handling
Reagent Tray 2 trays, 56 positions each, refrigerated
Reagent Capacity Onboard 52 colorimetric methods
Dispensing System 2 probes with Liquid Level Sensing and Liquid Surface Verification
Reagent Wedges 20, 40, 70 mL bar-code-labeled wedges
Reagent Inventory Management Tracks tests remaining, lot number, onboard stability, and expiration date
Reagent Dilutions Capability to dilute concentrated reagents onboard
Onboard Stability Up to 60 days
Open System Capability
Channels 200 channels, includes user-defined applications
3rd Party Applications Varies by country, can be configured on system
ISE
ISE Indirect simultaneous measurement of Na+, K+, Cl –
Sample Volume 22 μL for all three tests
Priming Automatic priming cycle
Electrode Shelf Life 30,000 samples or 3 months
Throughput Rate 600 tests/hour; 200 tubes/hour
Calibration/QC
Calibration Interval Up to 60 days, tracked by software
Auto-calibration/Auto-QC User-defined time interval or with new reagent container
View Calibration/QC Data Graphical display of calibration curves and QC; Siemens Remote Service QC monitoring
Calibration/Control Tray 61 refrigerated positions for calibrators, controls, and diluents
Data Management
Operating Computer Windows® XP, 1 GB RAM, touch screen monitor 19”
System Documentation Operator manual and assay sheets online
Host Interface RS 232C bidirectional
Data Storage 70,000 patient tests
Onboard Maintenance Logs Schedule and monitor routine maintenance activities via software
Host Query ASCII; system requests work order or batch of work orders from host
General Specifications
Power Requirements 200/220/230/240 V +/– 10%, 20 A, 50/60 Hz, 3 KVA
Water Requirements Deionized water from pressurized water (10-30 psi/169-207 KPa) Average Consumption: 30 liters per hour
Drain Requirements Minimum of 40 liters per hour
Dimensions ADVIA 1800: 44.6 (h) x 58.3 (w) x 34.5 (d) in, 113.3 (h) x 148.0 (w) x 87.6 (d) cm
Universal Rack Handler: 37 (h) x 29 (w) x 41 (d) in, 86.4 (h) x 73.7 (w) x 104.1 (d) cm
Weight ADVIA 1800: 1,323 lbs (600 kg)
Universal Rack Handler: 178 lbs (81 kg)
Compliance UL, cUL, CE
Noise Specifications Open cover less than 70 db
Average Heat Output 4,299 BTU/hour @ 50 Hz, 3,023 BTU/hour @ 60 Hz
Operating Temperature Range 18°-30°C/64°-86°F
Ambient Humidity 40%-70%